Emergency Stop And Automatic Processing After Stopping - Olympus OER-Pro Operation Manual

Endoscope reprocessor
Hide thumbs Also See for OER-Pro:
Table of Contents

Advertisement

Chapter 6 Other Functions
6.10 Emergency stop and automatic processing after
stopping
178
The equipment can be stopped at any time during operation by pressing the
STOP/RESET button on the main control panel. The main control panel will
display error code [E00] at this time. When the equipment is stopped, it identifies
the situation in which it was stopped and automatically performs the required
processing operations (including draining of the reprocessing basin), after which
it stops completely.
When a process is interrupted, be sure to execute it again
from the beginning. An endoscope that has not been
subjected to a complete, error-free process is not safe for
patient use. Otherwise, the endoscope reprocessing may be
insufficient.
If the process is stopped because of a problem with the
equipment, contact Olympus.
Do not set the power switch to OFF to stop the equipment
during operation. Otherwise, fluid will remain in the
reprocessing basin and the equipment may malfunction.
Do not set the power switch to OFF during automatic
processing. Otherwise, fluid will remain in the reprocessing
basin without automatic draining and air purge, and the
equipment may malfunction.
There are cases in which automatic processing can be interrupted and those in
which it cannot. Whether or not automatic processing can be interrupted should
be determined according to the displayed error code.
When the displayed error code remains lit
Pressing the STOP/RESET can interrupt the automatic
processing.
When the displayed error code is blinking
Automatic Processing cannot be interrupted.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual