 Draining water from the water filter housing
1.
Make sure that the power switch is ON.
2.
Close the lid by pushing it until it clicks.
3.
Open the front door of the device.
4.
Place the container with a capacity of 2 L or more in front of the device.
5.
Put the tube-side ends of the two filter tubes in the container placed above.
6.
Insert the connector ends of the two filter tubes into the connector below the
water filter housing and the connector above the water filter housing until
they click. Water will start to flow from the tube connected to the connector
below the water filter housing.
7.
Press the FUNC SEL button on the subcontrol panel to select "AIR
PURGE".
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Replace the water filter in a clean environment. Do not touch
the inner side of the water filter or allow dust to get in it.
After replacing the water filter, be sure to perform the
operation described in Section 7.3, "Disinfecting the water
supply piping" on page 202 to prevent the growth of
water-borne microorganisms. Failure to perform this
operation could result in contamination of the device piping
and ineffective reprocessing of the endoscope.
Always be sure to attach the water filter. Otherwise,
water-borne microorganisms and particulates in the water
may contaminate the device piping and prevent effective
reprocessing of the endoscope.
If air gets in the water filter housing, it may extend the
process time. In case of an irregularity such as extension of
the process time or lack of water supply, drain the air as
described in "Draining air in the water filter housing" on
page 199.
Be careful not to drop the removed water filter housing to
avoid damaging the connector below the water filter housing.
Make sure that the O-ring at the head of the water filter
housing is free of abnormalities such as cracks, breaks, rips,
scratches, or stains. Water leakage may result if the O-ring is
not attached or is abnormal.
Chapter 7 Routine Maintenance