Chapter 6 Other Functions
174
 If error code [E93] is displayed while alcohol is still present
in the alcohol tank:
If error code [E93] is displayed and the equipment is stopped while alcohol is still
present in the alcohol tank, perform the following steps.
The alcohol used with the equipment must be 70% ethyl
alcohol or isopropyl alcohol. Using any other kind of alcohol
may result in malfunction of the equipment or the endoscope,
difficulty drying the endoscope, fire hazard, or a hazard due
to toxic vapor emitted from the alcohol.
Alcohol is flammable and should be handled with extra care.
Remove the alcohol in the alcohol tank and replace it with
new alcohol at least once a week. Otherwise, the alcohol in
the alcohol tank may degrade.
Before handling the alcohol, read the precautions carefully
and use it as instructed.
1.
Connect the connecting tube to the connector at the rear center of the
reprocessing basin by pushing the tube's connector until it clicks.
2.
Close the lid and make sure that the power switch is ON.
3.
Connect the provided syringe and tube.
Figure 6.24
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual