Chapter 7 Routine Maintenance

Setting up the disinfectant solution

236
9.
Hold the lock levers and slowly disconnect the tube (see Figure 7.69).
Figure 7.69
10.
Rinse the disinfectant removal tube thoroughly under running water, dry it
completely, and store it in a clean place.
11.
Wipe the disinfectant removal port with the clean cloth and attach the rubber
cap to it.
12.
Close the front door.
The front door cannot be closed unless the rubber cap is
attached.
Before handling the disinfectant solution, read the
precautions carefully and use as instructed. It is especially
important to know what to do if the disinfectant solution
comes in contact with your skin.
Always use disinfectant that has been validated by Olympus.
High-level disinfectants that are not validated by Olympus for
use in the OER-Pro may be unsafe and ineffective due to
improper dilution, incorrect contact time and temperature,
excessive foaming, inadequate rinsing, and therefore may
compromise patient safety. Use of a high-level disinfectant
that has not been validated by Olympus may also damage
internal OER-Pro components (e.g., seals, valves, etc.) and
the endoscopes being reprocessed.