Chapter 3 Inspection Before Use
58
9.
Press the pin on the center of the metal connector of the leak test air tube to
confirm air supply.
Figure 3.23
10.
Disconnect the leak test air tube from the reprocessing basin's leak test
connector.
11.
With another leak test connector (black), confirm air supply by repeating
from Step 7 to 10.
Figure 3.24
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Pin