Emc Information - Olympus OER-Pro Operation Manual

Endoscope reprocessor
Hide thumbs Also See for OER-Pro:
Table of Contents

Advertisement

Appendix

EMC information

Emission
standard
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Main terminal
conducted
emissions
CISPR 11
Harmonic
Not applicable
emissions
IEC 61000-3-2
Voltage
Not applicable
fluctuations/flicker
emissions
IEC 61000-3-3
294
This model is intended for use in the electromagnetic environments specified
below. The user and medical staff should ensure that it is used only in these
environments.
 Magnetic emission compliance information and
recommended electromagnetic environments
Compliance
This equipment uses RF (radio frequency) energy only for its internal
function. Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
This equipment's RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
Power supply specification of this equipment is less than 220VAC, and this
equipment can be exempt from requirements of IEC 61000-3-2.
Power supply specification of this equipment is less than 220 VAC, and this
equipment is exempt from requirements of IEC 61000-3-3.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Guidance
Olympus confirmed that the level of this equipment's
harmonic emissions is low.