Appendix

Shipping environment

Operating environment

290
Shipping
Ambient temperatures –47 to +60C (–52 to +140F)
environment
Relative humidity
Atmospheric pressure
Operating
Ambient temperatures 10 – 40C (50 – 104F)
environment
Relative humidity
Elevation
Designed for use
Water supply flow
Water supply pressure 0.1 – 0.5 MPa (include water hammer)
Water supply
temperature
Water type
Water hardness
1 The recommended value is in reference to AAMI TIR34:2007 "Water for
the reprocessing of medical devices".
Purified water is water that has been produced through the
methods of either reverse osmosis (RO), deionization (DI),
distillation or other methods that meet USP standards to
remove impurities.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
10 – 90
700 – 1060 hPa
30 – 85
3000 meters (maximum)
Indoors
17 l/min. or more when the water faucet is
fully open
Max. 28C
Potable water or Softened water
(Purified water cannot be used)
400 ppm (maximum)
0 – 150 ppm (Recommended value)
1