Replacing The Fuse - Olympus OER-Pro Operation Manual

Endoscope reprocessor
Hide thumbs Also See for OER-Pro:
Table of Contents

Advertisement

Chapter 7 Routine Maintenance

7.14 Replacing the fuse

250
5.
Clean the mesh filter in running water using a soft brush.
Figure 7.86
6.
Place the mesh filter in the original position in the water supply connector.
Pay attention to the up-down orientation.
Be sure to install the mesh filter in the water supply hose
connector. Otherwise, dirt and foreign matter in the water
may enter the equipment and cause it to malfunction.
7.
Attach the connection ring of the water supply hose in the original position
on the equipment.
Replace the fuse as described in "If the lamp in the power switch does not turn
ON" on page 77.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual