Olympus OER-Pro Operation Manual page 213

Endoscope reprocessor
Hide thumbs Also See for OER-Pro:
Table of Contents

Advertisement

14.
Press the FUNC START button on the subcontrol panel again.
Figure 7.28
15.
Disconnect the filter tube as soon as disinfectant solution starts to flow from
it continuously.
Figure 7.29
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
If the temperature of the disinfectant solution is below 20C
(68F), it is heated to 20C (68F). During heating, the
TIME/CODE display blinks without counting down. It restarts
countdown after completion of heating.
Chapter 7 Routine Maintenance
207