18.
When the remaining time displayed on the main control panel reaches [00],
the buzzer beeps and the TIME/CODE display shows a [
as shown below.
Figure 7.32
19.
With the tube-side end of the filter tube held in place in the container, insert
the connector end of the filter tube into the connector above the water filter
housing until it clicks. Do not connect anything to the connector below the
water filter housing.
Figure 7.33
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Chapter 7 Routine Maintenance
] mark rotating
209