Chapter 7 Routine Maintenance
206
8.
Press the FUNC START button on the subcontrol panel. The TIME/CODE
display on the main control panel shows the remaining time, which is
counted down every minute.
Figure 7.26
9.
When the reprocessing basin is filled with disinfectant solution, three buzzer
beeps are generated and the TIME/CODE display on the main control panel
blinks.
10.
Push the area marked "PUSH" on the front door to open it.
11.
Prepare a container with a capacity of 2 L or more and place it in front of the
device.
12.
Hold the tube-side end of the filter tube and put it in the container.
13.
Insert the connector end of the filter tube into the connector above the water
filter housing.
Figure 7.27
Do not connect a filter tube to the connector below the water
filter housing.