Endoscope connector of the
180 or older series endoscope
(first endoscope)
1
190 series endoscopes may not be available in some areas.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
 When loading a 180 or older series endoscope and a 190
series endoscope
When placing the endoscope connector of the second
endoscope, hold the endoscope connector of the first
endoscope to avoid contact. Otherwise, the endoscope
connector of the first endoscope may be damaged by the
contact with S-cord connector mount of the second
endoscope.
• Pattern 1
First endoscope
Second endoscope
As shown in Figure 4.22, put the 190 series endoscope
while slightly lifting the endoscope connector of the 180 or older series
endoscope to the left (direction of arrow 1).
1
Figure 4.22
• Pattern 2
First endoscope
Second endoscope
As shown in Figure 4.23, tilt the endoscope connector of the 180 or older series
endoscope in place while slightly lifting the 190 series endoscope
the left (direction of arrow 1).
Chapter 4 Reprocessing Operations
1
simultaneously
: 180 or older series endoscope
: 190 series endoscope
Put the endoscope connector of the
second endoscope in this area
Endoscope connector of the 190
1
series
S-cord connector mount
Place the universal cord on the
inside of the hook
: 190 series endoscope
: 180 or older series endoscope
1
1
connector in place
(second endoscope)
1
1
connector to
99