Chapter 4 Reprocessing Operations
1
190 series endoscopes may not be available in some areas.
90
Place only the valves and other specified endoscope
components in the washing case in the reprocessing basin. If
any object other than those specified is placed in the washing
case, reprocessing of the endoscope valves will not be
effective.
The injection adapter (MAJ-1235) cannot be reprocessed
with the equipment. If it is placed in the washing case,
reprocessing of this adapter will not be effective.
If the endoscope has a forceps elevator, raise the forceps
elevator to an angle of approximately 45 and set it so that
the back of the forceps elevator can be sufficiently cleaned
and disinfected.
When reprocessing Gyrus ACMI endoscopes, rotate the
irrigation/biopsy port to ensure a proper fit into the OER-Pro
basin. Follow the instructions outlined in the MAJ-1904
instruction manual for proper endoscope configuration,
otherwise reprocessing may be insufficient.
The biopsy valve should be opened before being placed in
the reprocessing basin. Also, other accessories that can be
disassembled should be disassembled before being placed
in the washing case. Otherwise, they may not be sufficiently
reprocessed.
If the distal end cap of an endoscope is removable, remove
the distal end cap before putting the endoscope in the
reprocessing basin. Otherwise, reprocessing may be
insufficient.
Do not clog the circulation port inside the reprocessing basin.
Otherwise, the liquid feed pressure on the endoscopes will
be decreased and reprocessing will be insufficient.
When placing endoscopes into the reprocessing basin,
ensure that the endoscope(s) is loaded in a manner that it will
be fully submerged in the disinfectant, i.e., All parts of the
endoscope are below in the index bar as instructed in this
chapter.
Do not use the white-colored conventional retaining rack
(GL830600) for reprocessing 190 series endoscopes
Otherwise, reprocessing may be insufficient.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual