Disinfecting The Alcohol Supply Line; Figure 7.92 - Olympus OER-Pro Operation Manual

Endoscope reprocessor
Hide thumbs Also See for OER-Pro:
Table of Contents

Advertisement

22.
Disconnect the tube from the detergent nozzle and close the lid.
23.
Rinse the reprocessing basin by performing the rinsing process on the
subcontrol panel.
 Disinfecting the alcohol supply line
Check Required items
Connector jigs
Two cup-shape containers with a capacity of about 200 ml, such as a beaker
Disinfectant solution: approximately 80 ml
Sterile water: more than 80 ml
Table 7.18
1.
Make sure that the power switch of the device is ON.
2.
Make sure that the water faucet is open.
3.
Step on the foot pedal to open the lid.
4.
Attach the connector jigs to the connectors on the reprocessing basin and
close the lid.

Figure 7.92

5.
Pull out the detergent/alcohol drawer.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Make sure to attach the connector jigs. Otherwise, the
disinfection of the alcohol supply line may not be effective.
After disinfecting the alcohol supply line, always rinse it
thoroughly. Otherwise, disinfectant solution may remain on
the endoscope following reprocessing and pose a patient
safety risk.
Chapter 7 Routine Maintenance
255