The following figure shows the general flow of the OER-Pro's reprocessing
steps. For reliable reprocessing, it is important to understand the cleaning and
disinfection procedures thoroughly before using the equipment.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
For any endoscope that requires sterilization, always be sure
to sterilize the endoscope as instructed in its instruction
manual after cleaning/disinfection in the OER-Pro.
Certain endoscopes cannot be reprocessed with this
equipment. Refer to the provided "List of compatible
Endoscopes/Connecting Tubes <OER-Pro>" to see which
endoscopes are compatible. Accessories (e.g., valves) that
can be reprocessed with this equipment are accessories of
endoscopes compatible with this equipment. To reprocess
accessories, be sure to put them in the washing case. Do not
attempt to reprocess an endoscope and its accessories that
are not designated as compatible with the OER-Pro or that
are modified by a third party repair company; not only will the
equipment be unable to function at optimal levels, the safety
of the patient and operator may be endangered and this
equipment and/or the endoscope may be damaged.
When reprocessing JF-140F, TJF-140F, TJF-160F,
TJF-160VF, and TJF-Q180V duodenoscopes using
OER-Pro, conduct precleaning and manual cleaning as
detailed in each duodenoscope's reprocessing manual.
Chapter 4 Reprocessing Operations
73