Olympus OER-Pro Operation Manual page 233

Endoscope reprocessor
Hide thumbs Also See for OER-Pro:
Table of Contents

Advertisement

1.
Make sure that the power switch is ON.
2.
Step on the foot pedal to open the lid.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
If disinfectant solution leaks out of the disinfectant removal
port when the rubber cap has been removed, immediately
reattach the rubber cap and follow the procedure in Section
8.1, "Troubleshooting guide" on page 274. If it does not stop
leaking, contact Olympus.
When handling the disinfectant solution, wear appropriate
personal protective equipment to prevent it from making
direct contact with your skin and to prevent excessive
inhalation of the vapor. The disinfectant solution and its vapor
may adversely affect the human body. If you get disinfectant
solution in your eyes, immediately rinse with a large quantity
of water and then call the doctor. Personal protective
equipment, such as eyewear, face mask, moisture-resistant
clothing, and chemical-resistant gloves that fit properly and
are long enough so that your skin is not exposed. All
personal protective equipment should be inspected before
use and replaced periodically before it is damaged.
Be sure to disconnect the disinfectant collection hose after
collecting the disinfectant solution. If the device is operated
with the disinfectant collection hose connected, the cleaning
fluid and disinfectant solution may leak out of the device.
Be sure to disconnect the drain connector except when
collecting the disinfectant solution or checking its strength.
Otherwise, disinfectant solution may leak and damage the
equipment and areas near the device.
If "DRAIN LCG" is selected and the FUNC START button is
pressed without connecting the disinfectant collection hose,
the buzzer repeats short beeps, disinfectant solution will fill
the reprocessing basin and error code [E72] is displayed. In
this case, treat it by following the procedure in Section 8.1,
"Troubleshooting guide" on page 274.
Chapter 7 Routine Maintenance
227