After disinfection, rinse is performed three times. Rinse conditions such as
number of rinsing performed and rinsing time cannot be changed. Time and
water amount of each rinse cycle are shown below. Time and water amount
value vary depending on endoscope models, number of endoscopes loaded and
water supply conditions.
Number of rinses
Time
Rinse water volume
Table 6.3
1.
Set the power switch to ON.
2.
Press the PROG button on the main control panel to select the program to
be set.
Figure 6.5
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
1
100 s
12.4 L
3.3 gallons
Program [1] is preset to the values (cleaning time,
disinfecting time and disinfectant solution heating) and the
settings of this program cannot be altered.
Program [2] and [3] allow the user to extend the cleaning
time.
The disinfectant solution temperature takes about 4 minutes
for each 1C increase.
If you want to change disinfectant solution from ready-to-use
disinfectant solution to concentrated disinfectant solution,
contact Olympus.
Chapter 6 Other Functions
2
110 s
14.9 L
14.9 L
3.9 gallons
3.9 gallons
3
130 s
159