Chapter 8 Troubleshooting and Repair
Error
Problem
Code
E81
The process
cannot be
properly
controlled.
E82
Internal
irregularity in
the device.
E83
The process
cannot be
properly
controlled.
E84
Malfunction of
the RFID
reader.
E91
ID read error.
E92
Exceeded leak
test time.
E93
No alcohol
remains.
E94
The equipment
is not printing.
282
Possible Causes
Internal problem
with the device.
Irregularity in the
electrical circuitry
inside the device.
Irregularity in the
electrical circuitry
inside the device.
Irregularity in the
electrical circuitry
inside the device.
• ID is not read.
• Multiple IDs are
read.
Exceeded test
time (10 minutes).
• The alcohol has
run out.
• Alcohol cannot
be supplied to
the device.
• Irregularity in
the alcohol
sensor.
• Clog in the
pump or piping.
• Printer paper
roll has run out.
• The release
lever is raised.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Remedial Actions
Contact Olympus.
Contact Olympus.
Contact Olympus.
Contact Olympus.
Press the STOP/RESET button, check that the ID detection
indicators are not lit on the main control panel, and read the IDs
again.
Perform leak test again.
• If error occurred during the reprocessing process, refer to
"When the error code [E93] is displayed during the alcohol
flush process" on page 117.
• If error occurred during the sub-independent alcohol flush
process, check the amount of alcohol in the tank. If alcohol is
still present in the alcohol tank, refer to "If the error code
[E93] is displayed again after replenishing alcohol and
restarting the alcohol flush process:" on page 120.
• Install the printer paper roll properly (follow the procedure in
Section 4.8, "Installation of the printer paper roll" in the
"Instructions-Installation Manual").
• Check that the release lever is set in the proper position.