Error Result Printout - Olympus OER-Pro Operation Manual

Endoscope reprocessor
Hide thumbs Also See for OER-Pro:
Table of Contents

Advertisement

 ERROR RESULT PRINTOUT
If setting of Error Result Printout is activated, the printer prints error results
automatically when error occurs during the reprocessing process.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
If an error occurs during the error processing, the printer
prints the error result that occurred during the reprocessing
process. Identify the error code that occurred during the
reprocessing process and perform remedial action according
to Section 8.1, "Troubleshooting guide" on page 274.
If the printout is not satisfactory and the error code is
displayed on the main control panel, prints the latest
reprocessing record by "PRINT LAST" to confirm whether the
reprocessing process is completed. Press the "FUNC SEL"
button on the control panel to select "PRINT LAST" and then
press the "START" button. If the printer prints the
reprocessing results, it shows that the previous reprocessing
process was successfully completed. If the printer prints the
error results, it shows that the previous reprocessing process
was interrupted by error indicated in the printed error results.
For the printed information about "Error Results", refer to
"PRINT FORMAT (Error Result)" on page 146.
Chapter 4 Reprocessing Operations