Replacing The Air Filter (Maj-823) - Olympus OER-Pro Operation Manual

Endoscope reprocessor
Hide thumbs Also See for OER-Pro:
Table of Contents

Advertisement

Chapter 7 Routine Maintenance
7.5

Replacing the air filter (MAJ-823)

216
17.
Close the lid.
18.
Attach the gas filter case to the dome part of the lid.
19.
Perform the operation described in Section 6.7, "Air purge" on page 169 to
remove water from inner piping of the equipment.
20.
Step on the foot pedal to open the lid and take the sampling tube out of the
reprocessing basin, wipe off any water using a piece of clean cloth, and
store it in a clean place.
Replace the air filter every month.
Check Required items
Air filter (MAJ-823)
Indelible pen or other writing implement
Table 7.9
1.
Open the front door of the device.
2.
Remove the old air filter by pushing the sleeves on the two connectors
toward the device.
Sleeves
Figure 7.45
The air filter is not installed on the device when it is shipped
from the factory.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents