Chapter 4 Reprocessing Operations
120
The PRINT LAST lamp on the subcontrol panel lights up at
the end of reprocessing. You can also print the reprocessing
result of completed reprocessing process by manual
operation of "FUNC START" button.
 If the error code [E93] is displayed again after replenishing
alcohol and restarting the alcohol flush process:
When error code [E93] is displayed again after restarting the reprocessing
process, check that there is no irregularity regarding the following points.
The tube connected to the alcohol tank is not kinked.
The connector of alcohol tank is attached firmly.
If any irregularity is found by inspection, correct these points and press start
button to restart the alcohol flush process.
If any irregularity is not observed during an inspection or the error code [E93] is
displayed again, internal problem with the equipment may be suspected. In this
case, follow the procedure below.
Check Required items
Syringe
Tube
Table 4.1
The alcohol used with the equipment must be 70% ethyl
alcohol or isopropyl alcohol. Using any other kind of alcohol
may result in malfunction of the equipment or the endoscope,
difficulty drying the endoscope, fire hazard, or a hazard due
to toxic vapor emitted from the alcohol.
Alcohol is flammable and should be handled with extra care.
Remove the alcohol in the alcohol tank and replace it with
new alcohol at least once a week. Otherwise, the alcohol in
the alcohol tank may degrade.
Before handling the alcohol, read the precautions carefully
and use it as instructed.