Chapter 6 Other Functions
6.2
Heating the disinfectant solution
156
•
The following temperatures are displayed if the current
disinfectant solution temperature is as described below. Any
temperature between 1C and 20C are displayed as the
actual, measured disinfectant temperature.
Temperature
1C (34F) or less
20C (68F) or more
Table 6.1
•
When the current disinfectant solution temperature is below
20C (68F), the temperature display blinks.
This section explains how to heat the disinfectant solution independently of the
reprocessing programs. The operation consists of heating disinfectant solution to
20C (68F) when the disinfectant solution temperature is below 20C (68F).
For details on how to incorporate the disinfectant solution heating in a
reprocessing program, see Section 6.3, "Setting the reprocessing programs" on
page 158.
When heating the disinfectant solution, disconnect the
connecting tubes from the equipment. Otherwise,
disinfectant solution may spray out of the connecting tubes,
and leak from the reprocessing basin.
•
When heating the disinfectant solution, disconnect the leak
test air tube from the equipment and take it out from the
reprocessing basin. Otherwise, disinfectant solution may get
inside the leak test air tube and the endoscope. This could
cause them to malfunction.
•
Before starting heating the disinfectant solution, always
confirm that there is no foreign material on the ventilation
openings on the gas filter cases. Blocking the ventilation not
only hinders deodorization but may also cause the device to
malfunction.
For details on how to check the disinfectant solution
temperature, see Section 6.1, "Confirming the disinfectant
solution temperature" on page 155.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Display
1C
20C