Olympus OER-Pro Operation Manual page 283

Endoscope reprocessor
Hide thumbs Also See for OER-Pro:
Table of Contents

Advertisement

Error
Problem
Code
E00
Process is
interrupted.
E01
Filling the basin
with water takes
too long (water
supply time is
beyond
maximum
setting).
E02
Cleaning fluid is
not discharged.
E04
Cleaning fluid
decreases
during the
cleaning
process.
E05
Cleaning fluid
level is too high.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
If the automatic printing function is activated and printout is
not satisfactory due to the error that occurred during the
automatic printing, error code will be displayed on the main
control panel. In this case, print the latest reprocessing
record by "PRINT LAST" to confirm whether the reprocessing
process is completed. Press the "FUNC SEL" button on the
control panel to select "PRINT LAST" and then press the
"START" button. If the printer prints the reprocessing results,
it shows that the previous reprocessing process was
successfully completed. If the printer prints the error results,
it shows that the previous reprocessing process was
interrupted by error indicated in the printed error results.
Possible Causes
STOP/RESET
Wait until automatic processing is finished. During the error
button is pressed
processing, the error code may either blink or light steadily. If
during process.
you want to interrupt automatic processing and discharge fluid
inside the equipment, perform air purge. If you want to rinse the
basin, perform rinsing.
• The water
• Open the water faucet all the way.
faucet is not
• Check if water is leaking from the water supply hose.
open enough.
• Replace the water filter.
• Water leakage
• Clean the mesh filter in the water supply hose connector.
or clogging in
the water supply
piping.
• Irregularity in
• Check the drain hose for improper installation.
the drain tubing.
• Clean the mesh filter in the drain port of the basin.
• Clogging of the
• Close the lid and perform air purge to remove water.
mesh filter in the
drain port of the
reprocessing
basin.
Internal problem
Contact Olympus.
with the
equipment.
Irregularity in the
Contact Olympus (if water supply will not stop, close the water
fluid level sensor.
faucet).
Chapter 8 Troubleshooting and Repair
Remedial Actions
277