Olympus OER-Pro Operation Manual page 289

Endoscope reprocessor
Hide thumbs Also See for OER-Pro:
Table of Contents

Advertisement

Error
Problem
Code
E95
No detergent
remains.
Problem
The odor of the disinfectant
solution is stronger than
expected.
Water leaks from the
equipment.
Fluid leaks from the
disinfectant removal port.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Possible Causes
• The detergent
has run out.
• The detergent
cannot be
supplied to the
equipment.
• Irregularity in
the detergent
sensor.
• Clog in the
pump or piping.
 Other errors and remedial action
Possible Causes
• The gas filter has
expired.
• Disinfectant solution is
leaking.
Improper installation of
water filter housing.
Internal problem with the
device.
Something is clogging the
disinfectant removal port.
Remedial Actions
Refer to "When the error code [E95] is displayed during the
reprocessing process" on page 125.
• Replace the gas filter as described in Section 7.1,
"Replacing the gas filter (MAJ-822)" on page 188. If
the problem persists after replacement, contact
Olympus.
• Check if disinfectant solution is leaking. If it is, do not
use the device and contact Olympus.
Stop the current process and reattach the water filter
housing as described in Section 7.2, "Replacing the
water filter (MAJ-824)" on page 192.
However, the STOP/RESET button may not work if
stopping the water might cause a hazardous situation,
for example during disinfectant solution replacement. In
this case, close the water faucet, tighten the water filter
housing, then open the water faucet again and continue
the process. If an error code is displayed, take the
appropriate remedial action for that error.
Close the water faucet, set the power switch to OFF,
disconnect the power cord plug from the wall mains
outlet and contact Olympus.
Connect the drain connector as described in to Section
3.10, "Checking the disinfectant solution concentration
level" on page 59, push the valve on the connector
several times so that the material stuck in the port is
removed. If leakage is still detected, attach the rubber
cap to the disinfectant solution drain port and contact
Olympus.
Chapter 8 Troubleshooting and Repair
Remedial Actions
283