Olympus OER-Pro Operation Manual page 143

Endoscope reprocessor
Hide thumbs Also See for OER-Pro:
Table of Contents

Advertisement

1.
At the end of the reprocessing process, make sure that "PRINT LAST" is
selected on the subcontrol panel.
2.
Press the FUNC START button on the subcontrol panel. The printer prints
the results of the last reprocessing process.
Figure 4.53
3.
The buzzer beeps and the paper feed stops when printing completes. Cut
the printed part of paper and ensure that information is printed correctly.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
The "PRINT LAST" selection on the subcontrol panel is
deselected when any button is pressed after the end of the
process. If this happens, press the FUNC SEL button on the
subcontrol panel to select "PRINT LAST".
If the printout is not satisfactory, printing can be restarted.
Press the FUNC SEL button on the subcontrol panel to select
"PRINT LAST" and then press the FUNC START button.
The same information can be printed several times until the
next reprocessing process is started.
For the printed information about "Reprocessing Result",
refer to "PRINT FORMAT (Reprocessing Result)" on
page 145.
Chapter 4 Reprocessing Operations