ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
When reprocessing an ultrasonic endoscope, always use the
specially designed retaining rack (MAJ-840: blue colored). If
the retaining rack (MAJ-1970: pink colored) is used for the
ultrasonic endoscope, reprocessing may be insufficient.
When placing an endoscope that requires a water-resistant
cap (EVIS or EXERA video endoscope, ultrasonic
endoscope, etc.) in the basin, be sure to attach the
water-resistant cap as described in the endoscope's
instruction manual. If you forget to attach the water-resistant
cap or if you attach the water-resistant cap when it's wet
inside, the endoscope may malfunction due to fluid invasion.
When setting each endoscope in the reprocessing basin,
confirm that it is not excessively scratched, which could result
in water leakage. If any irregularity is observed, do not place
the endoscope in the equipment and contact Olympus for
repair.
To prevent damage, make sure to set up the endoscopes so
that the distal ends will not fall out of the retaining rack or
come in direct contact with the wall of the reprocessing basin.
When reprocessing an ultrasonic endoscope, always use the
specially designed MAJ-840 retaining rack (optional).
Otherwise, the endoscope or reprocessor may be damaged.
When heating the disinfectant solution, disconnect the leak
test air tube from the equipment and remove it from the
reprocessing basin. Otherwise, disinfectant solution may get
inside the leak test air tube and the endoscope and cause
damage.
You may reprocess four auxiliary water tubes (MAJ-855)
simultaneously using two MAJ-1904s. Refer to "List of
Compatible Endoscopes/Connecting Tubes <OER-Pro>" for
other combinations including endoscopes.