Chapter 4 Reprocessing Operations
Setting position for
first endoscope
connector
1
190 series endoscopes may not be available in some areas.
96
 When loading a 190 series endoscope
endoscope
Place the endoscope connector over the black marker located on the left side of
the divider as shown in Figure 4.17.
Black marker
Divider
Figure 4.17
5.
Adjust the positions of the insertion tube and universal cord to minimize
overlapping. Also, adjust the positioning of the insertion tube's distal end by
turning the angulation control knobs (lever) on the control section. Again,
make sure that the universal cord is placed on the inside of the hook (see
Figure 4.14).
6.
When placing a second endoscope, repeat Steps 1 to 4 above. The control
section of the second endoscope should be placed to the left of the first
endoscope as shown in Figure 4.18.
Control section of the
second endoscope
Figure 4.18
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
1
as a first
Divider