Chapter 7 Routine Maintenance
208
16.
A jet of fluid exits the hole on the water supply piping disinfection hose
during the process. Make sure that the fluid hits the lid.
Figure 7.30
17.
Make sure that a jet of fluid is output from the water supply/circulation
nozzle to the dome of lid during the process.
Figure 7.31
Fluid jet from the water supply/circulation nozzle of this
process is gentler than other processes such as the
endoscope reprocessing process. As a result, fluid spreads
on less than half of dome part of the lid whereas the fluid jet
of other process spreads on entire dome part of the lid. This
difference is due to use of a different pump and therefore the
gentle fluid flow of this process is adequate.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual