Chapter 3 Inspection Before Use
3.1

Inspecting for fluid leaks

40
Confirm that water or fluid does not leak from the water supply piping, inside the
equipment, Drain hose connector, etc.
Do not continue using the equipment if it is leaking water.
Doing so may result in an electric shock or malfunction.
If water or fluid leaks from inside the equipment, close the
water faucet, set the power switch to OFF, unplug the power
cord, and contact Olympus.
1.
Slowly open the water faucet.
2.
Confirm that water is not leaking from where the water supply hose is
connected to the water faucet or equipment.
3.
Confirm that water or fluid is not leaking on the floor from inside the
equipment.
 If water leaks from the water supply hose connector
1.
Close the water faucet.
2.
Check the installation of the water supply devices by referring to Section
4.2, "Connection of the water supply hose" in "Instructions-Installation
Manual".
 If water leaks from the water filter housing
1.
Close the water faucet.
2.
Check the installation of the water filter housing by referring to Section 3.9,
"Installation of the water filter (MAJ-824)" in "Instructions-Installation
Manual".
 If water or fluid leaks from inside the equipment
1.
Close the water faucet.
2.
Set the power switch of the equipment to OFF.
3.
Unplug the power cord from the power outlet.
4.
Contact Olympus.