Chapter 7 Routine Maintenance
258
14.
Prepare a beaker or similar container and fill it with at least 80 ml of sterile
water.
15.
Put the cap in the container filled with the sterile water (see Figure 7.97).
16.
Rinse the alcohol supply line by performing the alcohol flushing process on
the subcontrol panel. Check that the sterile water in the container is
decreasing as the process progresses. The process ends automatically in
three minutes (see Figure 7.96).
17.
Attach the cap to the alcohol tank and close the detergent/alcohol drawer
(see Figure 7.97).
Figure 7.97
18.
Remove rinsing water from inside the line and fill the line with alcohol by
performing the alcohol flush process on the subcontrol panel. The process
ends automatically in three minutes (see Figure 7.96).
19.
Rinse the reprocessing basin by performing the rinsing process on the
subcontrol panel.
20.
Open the lid and disconnect the connector jigs.