Chapter 4 Reprocessing Operations
Loading the endoscope and their accessories into the
OER-Pro
4.4

Scope ID detection

84
 Flushing the interior of the forceps elevator
1.
Immerse the distal end in detergent solution, and operate the elevator
control lever in accordance with the reprocessing instructions provided in
the endoscope manual.
2.
With the forceps elevator raised, flush the interior of the forceps elevator in
accordance with the reprocessing instructions provided in the endoscope
manual.
 Cleaning the endoscope's accessories
Manually clean the reusable parts such as the air/water valve, suction valve and
biopsy valve, according to the instructions provided in the endoscope manual.
Carefully lift the endoscope out of the detergent solution, allowing excess fluid to
drain into the basin. Carry the endoscope to the reprocessing basin of the
OER-Pro. Place the endoscope in the reprocessing basin and connect the
required Connecting Tube(s) to the endoscope and put the valves in the washing
case in the center of the retaining rack according to the Instruction Manual of the
OER-Pro. Continue reprocessing, according to the Instruction Manual of the
OER-Pro.
To maintain an endoscope reprocessing log, this equipment is capable of
recognizing the individual scope ID that identifies the endoscope being
reprocessed. The RFID function that enables this operation is deactivated at the
factory. If you want to use this function, contact Olympus. When the RFID
function is disabled, neither the scope nor user ID is detected. If you are not
using the scope ID function, you may skip this section and the following section
and go to Section 4.6, "Endoscope setup" on page 89.
This equipment can be set up to use the RFID function.
Please be aware that the radio waves emitted from the RFID
reader of the equipment may cause medical devices such as
pacemakers to malfunction. If any interference with the
equipment is observed, immediately move away from the
RFID reader or set the power switch to OFF. Call your doctor
if you do not begin to feel better.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual