Inspection Of The Amount Of Alcohol - Olympus OER-Pro Operation Manual

Endoscope reprocessor
Hide thumbs Also See for OER-Pro:
Table of Contents

Advertisement

Inspection of the amount of alcohol

Check the ALCOHOL indicator on the detergent/alcohol drawer of the equipment
to confirm that alcohol is present. If the amount of alcohol has decreased, add
alcohol as described in "Addition of alcohol" before the level reaches "MIN.".
Figure 3.10
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
If alcohol flushing is initiated without alcohol in the tank, the
error code [E93] will be displayed and the process will be
stopped.
When reprocessing is performed with newly-installed
equipment or with equipment after Section 7.16, "Preparing
the reprocessor for long-term storage" on page 259 is
performed, reprocessing may stop with the error code [E95]
displayed even though there is enough detergent in the tank.
See page 174 ("If error code [E93] is displayed while alcohol
is still present in the alcohol tank:") to solve this problem.
If alcohol flushing is executed when no alcohol is present, the
equipment will stop and display an error code [E93].
Chapter 3 Inspection Before Use
49