Chapter 3 Inspection Before Use
62
Recommended volume for sampling of the disinfectant
solution is 20 ml or less. If too much disinfectant solution is
remove for sampling, over time the level of disinfectant
solution in the disinfectant tank will be reduced, and may
cause Error Code [E12] or [E14].
Figure 3.27
4.
While taking care not to inhale the disinfectant solution vapor, immerse the
test strip in the disinfectant solution in the container to check the
concentration. If the concentration is below the effective level, replace the
disinfectant solution as described in Section 7.12, "Replacing the
disinfectant solution" on page 225.
5.
Place the prepared cloth under the drain connector, hold the lock lever and
slowly disconnect the connector. Wipe off any disinfectant solution if it leaks.
The disinfectant should not be used beyond the duration of
its reuse life, even if tests taken during that time show that it
is at or above its minimum recommended concentration.
A disinfectant may fall below its minimum recommended
concentration before the end of its reuse life.
6.
Wipe the disinfectant removal port with a clean cloth and put the rubber cap
back on. Rinse the disinfectant drain connector thoroughly in running water,
dry it thoroughly and store in a clean place.
7.
Close the front door.
The front door cannot be closed unless the rubber cap is
attached.