13.
Open the front door and remove the rubber cap from the disinfectant
removal port.
Figure 7.61
14.
Put the tube-side end of the disinfectant removal tube in the container
(small), and connect the disinfectant removal tube to the disinfectant
removal port to drain the small amount of residual disinfectant solution from
the disinfectant solution tank (see Figure 7.62).
Figure 7.62
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Chapter 7 Routine Maintenance
Disinfectant
removal port
Rubber cap
231