Chapter 7 Routine Maintenance

7.10 Checking cassette cutters

7.11 Cleaning the disinfectant bottle drawer

224
7.
Turn the connectors to correct the orientations of the tubes as shown below.
Connectors
Figure 7.55
8.
Close the detergent/alcohol drawer.
This procedure is applied only when using concentrated
disinfectant prepared by cassette loading. When using
ready-to-use disinfectant prepared by top loading, the drawer
is locked.
Inspect the cassette cutters when replacing the disinfectant solution. For details,
see Section 7.12, "Replacing the disinfectant solution".
This procedure is applied only when using concentrated
disinfectant prepared by cassette loading. When using
ready-to-use disinfectant prepared by top loading, the drawer
is locked.
It is necessary to clean the disinfectant bottle drawer if the disinfectant bottle
drawer is dirty or any disinfectant solution spills into the drawer. The disinfectant
solution bottle drawer can be opened only when replacing the concentrated
disinfectant solution. For details, see Section 7.12, "Replacing the disinfectant
solution".