Chapter 4 Reprocessing Operations
88
Electromagnetic interference with other devices may shorten
the communications distance of the designated ID tag and
cause signals to become unreadable. Try to take mitigation
measures such as keeping other radio frequency emitting or
other energy emitting devices, such as electrosurgical units,
away from this equipment.
The START button on the main control panel will not function
until the user ID card has been recognized.
Do not leave a user ID card near the RFID reader.
You can print out the user name read from the user ID card
with the printer.
For addition or reissue of a user ID card, contact Olympus.
If the equipment fails to detect the user ID, apply the provided
user ID master card to the RFID reader so that the equipment
recognizes the user ID as the master ID. If a problem with the
scope ID is observed, contact Olympus.
1.
Hold the reprocessing operator's user ID card parallel to the RFID reader
and scan the card with the reader until the equipment generates a short
beep.
Figure 4.9
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual