3.10 Checking the disinfectant solution concentration
level
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
12.
After removing the leak test air tube from the reprocessing basin, close the
lid to drain the water and finish leak test. At this time, the buzzer beeps and
the main control panel displays [- -] to indicate the end of testing.
Figure 3.25
13.
Wipe off any water on the outer surface of the leak test air tube with a clean
cloth, and store it in a clean place.
Prior to reprocessing, check the concentration of the disinfectant solution using a
test strip to verify that the concentration of disinfectant solution meets the
minimum recommended concentration. Replace the disinfectant solution when it
fails to meet the minimum recommended concentration or beyond the specified
use life.
Check Required items
FDA-cleared chemical indicator (test strip)
Drain connector (should be dry)
Cup-shape container with a capacity of about 200 ml, such as a beaker
(should be dry)
Clean cloth
Table 3.8
Chapter 3 Inspection Before Use
59