Chapter 7 Routine Maintenance
230
10.
When draining is stopped, the long buzzer beeps, and the TIME/CODE
display on the main control panel shows [- -] indicating the end of the
process.
Figure 7.59
11.
Pull the sleeve on the connector of the disinfectant collection hose to
disconnect it from the disinfectant solution nozzle. Be sure to drain the
disinfectant solution that remains in the disinfectant collection hose into the
container (large).
Figure 7.60
12.
Rinse both the outside and inside of the disinfectant collection hose
thoroughly in running water, dry it thoroughly, and store it in a clean place.