ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
If you are reprocessing two endoscopes in the OER-Pro, and
a problem is observed with the connecting tubes on one of
the endoscopes, correct the problem and then reprocess
both endoscopes again, starting from the beginning.
Disconnect the connecting tubes from the connectors on the
equipment whenever the tubes are not used for
reprocessing. If reprocessing is performed while the
unnecessary tubes are connected, the effectiveness of
cleaning/disinfection may be reduced.
When closing the lid, be careful not to get the connecting
tube caught between the reprocessing basin and lid and
make sure the endoscopes and the washing case are not
touching the lid. Close the lid after ensuring that the cover of
the washing case is closed. If the endoscopes and the
washing case are touching the lid, adjust their positions and
close the lid. If the lid is closed while contacting the
endoscopes or accessories such as the washing case and
the connecting tubes, the endoscopes, the accessories and
the equipment may get damaged or water leakage may
result.
To confirm the correct connecting tubes are attached to the
endoscope, always refer to the instruction manuals of
endoscope or the latest "List of Compatible
Endoscopes/Connecting Tubes <OER-Pro>" that is marked
"For pink colored retaining rack (MAJ-1970)". Using incorrect
connecting tubes may result in ineffective reprocessing of
endoscopes. If you do not have the latest "List of Compatible
Endoscopes/Connecting Tubes <OER-Pro>" that is marked
"For pink colored retaining rack (MAJ-1970)", contact
Olympus.
Disconnect the leak test air tube from the connectors on the
equipment whenever the tubes are not used for leak test. If
reprocessing is performed while the unnecessary tubes are
connected, fluid may get inside the leak test air tube and
could cause it to malfunction. The fluid inside the tube may
also damage the endoscope when the leak test is performed
next time.
Chapter 4 Reprocessing Operations
109