Treatment after automatic processing has completed
Remove the cause of the equipment stoppage, and start the process again from
the beginning.
Processing after automatic processing is interrupted
Remove the cause of the interruption. If water or cleaning fluid remains in the
reprocessing basin, clean the basin with the following procedure. Also, the
stopped process must be started again from the beginning.
1.
Press the FUNC SEL button on the subcontrol panel to select "AIR
PURGE".
2.
Press the FUNC START button on the subcontrol panel to drain the
remaining water or fluid.
3.
When air purge is finished, press the FUNC SEL button to select "RINSE".
4.
Press the FUNC START button on the subcontrol panel to start the rinsing
process.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Chapter 6 Other Functions
179