Olympus OER-Pro Operation Manual page 149

Endoscope reprocessor
Hide thumbs Also See for OER-Pro:
Table of Contents

Advertisement

Error occurred
Print Option
during reprocessing
AUTOMATIC
PRINTING
PRINT LAST
PRINT TODAY
PRINT ALL
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
The print function can print completed reprocessing cycles or
error codes that occur during the reprocessing cycle. It
cannot print the record of completed/incomplete
sub-independent processes and the record of error occurred
during the standby operation. For details on the print function
of error result, see Table 4.12.
Error occurred when
starting
reprocessing cycle
cycle
(ex. E12, E31, E91)
Print option can print out the
subject error result
Table 4.12
1.
Press the FUNC SEL button on the subcontrol panel so that both the PRINT
LAST and PRINT TODAY indicators light up.
2.
Press the FUNC START button on the subcontrol panel. The printer prints
the results of all of the reprocessing events stored in memory in reverse
order, beginning with the most recent.
Figure 4.56
3.
The buzzer beeps and paper feed stops when printing is finished. Cut the
printed part of paper and ensure that the information is printed correctly.
Error occurred when
in standby
Print option cannot print out the
subject error result
Chapter 4 Reprocessing Operations
Error occurred
during
sub-independent
process
143