Error Codes And Remedial Actions - Olympus OER-Pro Operation Manual

Endoscope reprocessor
Hide thumbs Also See for OER-Pro:
Table of Contents

Advertisement

Chapter 8 Troubleshooting and Repair
276
If a rinse is performed while an error is actively being
displayed and disinfectant solution is still in the reprocessing
basin, the rinse process will drain the disinfectant out of the
device. To prevent this, refer to "Error codes and remedial
actions" for instructions for collecting the disinfectant.
Do not press the power switch OFF when an error code is
displayed. Doing so may result in malfunction. The device will
automatically start a process to address the error code. (The
error code will blink during automatic processing.) After the
error code starts blinking, follow the instructions in "Error
codes and remedial actions" below.
 Error codes and remedial actions
1.
If the equipment detects irregularity, buzzer beeps for three seconds and the
error code will be displayed on the main control panel (ex. E95 will be
displayed when the detergent has run out). While the error code on the main
panel blinks, the equipment executes the automatic processing. In this case,
wait few minutes until the blinking of error code changes to a steady light.
Figure 8.1
2.
When the automatic processing is completed, the buzzer beeps for three
seconds. The error code remains on the main control panel and the buzzer
beeps intermittently until the error code is cleared. Identify the error code
and perform remedial actions according to the instruction described in the
table below.
The error code displayed on the main control panel can be
cleared only by pressing the STOP/RESET button.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual