Chapter 4 Reprocessing Operations
94
4.
Wind the universal cord counterclockwise around the inside section of the
retaining rack by placing it under the hooks so that it will not move upward
(see Figure 4.14). If the scope ID tag is attached to the endoscope, move
the scope ID tag toward the endoscope connector. Straighten the ID tag's
band if it is twisted.
Figure 4.14
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Place on the inner side
of the hook