Disinfecting The Internal Detergent/Alcohol Lines; Draining And Using The Disinfectant Solution From The Equipment; Disinfecting Of The Detergent Supply Line - Olympus OER-Pro Operation Manual

Endoscope reprocessor
Hide thumbs Also See for OER-Pro:
Table of Contents

Advertisement

7.15 Disinfecting the internal detergent/alcohol lines

Draining and using the disinfectant solution from the
equipment
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
When handling the disinfectant solution and
detergent/alcohol, carefully read the precautions for its use to
fully understand the given information and use as instructed.
Particular understanding is required for measures to be taken
in case the disinfectant solution comes into contact with your
skin.
When handling the disinfectant solution and
detergent/alcohol, wear appropriate personal protective
equipment to avoid direct contact with your skin or excessive
inhalation of its vapor. The disinfectant solution and its vapor
may affect the human body. Personal protective equipment,
such as eyewear, face mask, moisture-resistant clothing, and
chemical-resistant gloves that fit properly and are long
enough so that your skin is not exposed. All personal
protective equipment should be inspected before use and
replaced periodically before it is damaged.
To prevent spills, keep the detergent/alcohol tanks upright.
Collect the disinfectant solution as described in Section 3.10, "Checking the
disinfectant solution concentration level" on page 59. Always check the
disinfectant solution concentration and use it to disinfect detergent/alcohol
supply pipe only when it's effective.
 Disinfecting of the detergent supply line
Check Required items
Syringe and tube
Two Cup-shape container with a capacity of about 200 ml, such as a beaker
Disinfectant solution: approximately 30 ml
Sterile water: more than 100 ml
Table 7.17
Chapter 7 Routine Maintenance
251

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents