Table of Contents

Advertisement

4.
Turn the cap on the alcohol inlet in the direction shown to remove the cap.
Cap
Figure 3.13
5.
Pour alcohol until it is level with the line inside the alcohol tank. Be careful
not to spill any. If any alcohol spills from the tank, wipe it off with a clean
cloth. After adding the alcohol, replace the cap on the alcohol tank.
Figure 3.14
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Do not add alcohol while the tank is in the detergent/alcohol
drawer. If alcohol is spilled on the tray, it could damage the
equipment.
Do not tilt the alcohol tank while there is still alcohol inside.
Otherwise, the alcohol may spill.
The amount of alcohol required to fill the tank up to the level
line is about 500 ml (approximately 16 ounces, enough for
about 10 alcohol flushes).
Chapter 3 Inspection Before Use
Alcohol inlet
Line
51