7.12 Replacing the disinfectant solution

ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
When the disinfectant solution in the device is no longer effective, or beyond the
specified use life, drain the disinfectant solution completely and replace with
fresh disinfectant solution. Expired disinfectant solution should be treated as
directed in the documents supplied with the disinfectant solution.
Before handling the disinfectant solution, read the
precautions carefully and use it as instructed. Be sure that
you fully understand what measures need to be taken if you
get any disinfectant solution on your skin.
Always use a disinfectant that has been validated by
Olympus. High-level disinfectants that are not validated by
Olympus for use in the OER-Pro may be unsafe and
ineffective due to improper dilution, incorrect contact time
and temperature, excessive foaming, inadequate rinsing, and
therefore may compromise patient safety. Use of a high-level
disinfectant that has not been validated by Olympus may also
damage internal OER-Pro components (e.g., seals, valves,
etc.) and the endoscopes being reprocessed.
When handling the disinfectant solution, wear appropriate
personal protective equipment to prevent direct contact with
your skin or excessive inhalation of the vapor. The
disinfectant solution and its vapor may adversely affect the
human body. If you get disinfectant solution in your eyes,
immediately rinse with a large quantity of water and then call
the doctor. Personal protective equipment, such as eyewear,
face mask, moisture-resistant clothing, and
chemical-resistant gloves that fit properly and are long
enough so that your skin is not exposed. All personal
protective equipment should be inspected before use and
replaced periodically before it is damaged.
Expired disinfectant solution should be treated in accordance
with the instructions supplied with the disinfectant solution. It
is recommended to treat the waste fluid properly and to
dispose of it according to local wastewater standards defined
by law, or temporarily collect and store the waste fluid and
have it treated by a waste disposal firm.
Chapter 7 Routine Maintenance