Print Format (Reprocessing Result) - Olympus OER-Pro Operation Manual

Endoscope reprocessor
Hide thumbs Also See for OER-Pro:
Table of Contents

Advertisement

PRINT FORMAT (Reprocessing Result)

P r o c e s s
S t a r t e d :
2 0 1 2 / 0 4 / 0 1
1 4 : 3 0
P r o c e s s
C o m p l e t e d :
2 0 1 2 / 0 4 / 0 1
1 5 : 0 0
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Serial number of this equipment
Total accumulated process count
Year/Month/Day and Hour/Minute
of the start of process
Year/Month/Day and Hour/Minute
of the end of process
Setting details
• Wash
• Dis
• Temp
• Leak
• Alc
Serial number of scope
Model name of scope
Serial number of scope (when two scopes were reprocessed)
Model name of scope (when two scopes were reprocessed)
User number
User name
User check columns
The check results of precleaning, connecting tube
connection conditions and disinfection
concentration test can be entered by the user.
Figure 4.57
If the RFID function has not been enabled at the factory, the
columns for the serial numbers of endoscopes, model names
of endoscopes, user number, and user name will be left
blank.
Chapter 4 Reprocessing Operations
Cleaning time (minutes)
Disinfecting time (minutes)
Disinfectant solution temperature (C)
"" when leak test is executed
"" when leak test is executed
"" when alcohol flushing is executed
145