Chapter 4 Reprocessing Operations
116
5.
When reprocessing is finished, the buzzer beeps and the TIME/CODE
display shows [- -] indicating that the process has ended.
Figure 4.38
When the process is started, the standard time required to
complete the process will blink on the display.
The minimum required time for program [1] is about
26 minutes (when using Olympus-validated concentrated
disinfectant solution) or 29 minutes (when using
Olympus-validated ready-to-use disinfectant solution).
This varies depending on the models and number of
endoscopes, ambient temperature of the disinfectant
solution, and water supply volume.
To display the correct time information, the equipment
measures the water supply time of the first process,
calculates the required process time automatically, and
corrects the displayed value automatically at the beginning of
cleaning (or end of water supply) during the reprocessing
process.
The equipment may also update the displayed time
information during the rinsing process.
If the "AUTOMATIC PRINTING" is activated, the printer
automatically prints the reprocessing result of completed
reprocessing cycle. For more details on the "AUTOMATIC
PRINTING", refer to Section 4.11, "Printing of the
reprocessing records" on page 130.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual