Coordinates (With Distortion Correction Enabled - Device Dependend) - Baumer VeriSens ID Series Technical Documentation Manual

Vision sensor
Hide thumbs Also See for VeriSens ID Series:
Table of Contents

Advertisement

9.1.3 Coordinates (with distortion correction enabled – device dependend)
When distortion correction is enabled there is an option to convert the internal image coordinate system
(which uses pixel as its unit) to a user-defined coordinate system.
This requires the real coordinates in the image to be calibrated to a world coordinate system with a
calibration template or manually.
NOTE
A calibration template is required for calibration using the Point, Target grid and SmartGrid
methods.
After successful installation, the calibration templates can be found in the directory:
<installation path>\AppSuite\calibration
on your PC.
NOTE
The Image recording tab has to be used to activate Enable distortion correction... to allow
configurations to be entered here.
v2.11.0-B4