Voltage Transient Example
The table and associated figure below illustrates the operation of a transient. The blue line
represents the RMS value of the phase A output voltage.
Volt AC
120.0
110.0
100.0
90.0
80.0
70.0
60.0
50.0
40.0
20.0
0.0
0.0
To generate this voltage versus time output sequence, the following transient list entries are
required. (Ramp and dwell times shown in milliseconds using STEP mode.)
#
1
2
3
4
5
6
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS) • All Rights Reserved • No reproduction without written authorization from PPS.
LMX Series Power Source Operation Manual
#2
Step #1
Step #3
1.0
2.0
3.0

Figure 6-9: Voltage Transient Example 1

Ramp
1000
0.2
0.2
1000
500
0.5

Table 6-8: Voltage Transient List for Example 1

LMX SERIES™ OPERATION MANUAL
SECTION 6:
Step #4
115Vac
4.0
5.0
Freq
Volt
400
100.00
400
0.00
400
100.00
400
115.00
400
80.00
400
115.00
Front Panel Operation
Step #5
Step #6
6.0
7.0
Time à
Dwell
1000
100
900
1000
1000
1.00
Page 122 of 463